Go to content
UR Home

Commissioning of volumetric modulated arc therapy (VMAT) in a dual-vendor environment

Dobler, Barbara and Groeger, Christian and Treutwein, Marius and Alvarez Moret, Judith and Goetzfried, Thomas and Weidner, Karin and Härtl, Petra and Koelbl, Oliver (2011) Commissioning of volumetric modulated arc therapy (VMAT) in a dual-vendor environment. Radiotherapy and Oncology (99), pp. 86-89.

[img]
Preview
PDF - Accepted Version
Volltext
Download (228kB)

at publisher (via DOI)

at PubMed

Other URL: http://dx.doi.org/10.1016/j.radonc.2011.01.024


Abstract

Methods and results for commissioning of the complete VMAT delivery chain are presented for the combination of Nucletron's Oncentra MasterPlan® v3.3 with Elekta's Mosaiq® v1.6 and SynergyS® linac. VMAT specific linac commissioning included determination of the size of the minimal dynamic leaf gap. Dosimetric validation of the complete treatment chain was performed using a 2D-ionization-chamber-array and showed excellent dosimetric results.


Export bibliographical data



Item Type:Article
Date:2011
Institutions:Medicine > Lehrstuhl für Strahlentherapie
Identification Number:
ValueType
10.1016/j.radonc.2011.01.024DOI
21458090PubMed ID
Keywords:Volumetric modulated arc therapy (VMAT); Commissioning; Clinical implementation
Dewey Decimal Classification:600 Technology > 610 Medical sciences Medicine
Status:Published
Refereed:Yes, this version has been refereed
Created at the University of Regensburg:Yes
Deposited On:06 Feb 2012 13:28
Last Modified:03 Jul 2015 11:18
Item ID:23390
Owner Only: item control page

Downloads

Downloads per month over past year

  1. Homepage UR

University Library

Publication Server

Contact person
Gernot Deinzer

Phone +49 941 943-2759
Contact