Dobler, Barbara and Groeger, Christian and Treutwein, Marius and Alvarez Moret, Judith and Goetzfried, Thomas and Weidner, Karin and Härtl, Petra and Koelbl, Oliver (2011) Commissioning of volumetric modulated arc therapy (VMAT) in a dual-vendor environment. Radiotherapy and Oncology (99), pp. 86-89.

PDF - Accepted Version
Download (228kB)

at publisher (via DOI)

at PubMed

Other URL: http://dx.doi.org/10.1016/j.radonc.2011.01.024


Methods and results for commissioning of the complete VMAT delivery chain are presented for the combination of Nucletron's Oncentra MasterPlan® v3.3 with Elekta's Mosaiq® v1.6 and SynergyS® linac. VMAT specific linac commissioning included determination of the size of the minimal dynamic leaf gap. Dosimetric validation of the complete treatment chain was performed using a 2D-ionization-chamber-array and showed excellent dosimetric results.