In vivo comparison of braided (Accero) and laser-cut intracranial stents (Acclino, Credo): evaluation of vessel responses at subacute and mid-term follow-up in a rabbit model
Mühl-Benninghaus, Ruben, Tomori, Toshiki, Krajewski, Stefanie, Dietrich, Philipp, Simgen, Andreas, Yilmaz, Umut, Brochhausen, Christoph, Kießling, Mara, Reith, Wolfgang and Cattaneo, Giorgio (2020) In vivo comparison of braided (Accero) and laser-cut intracranial stents (Acclino, Credo): evaluation of vessel responses at subacute and mid-term follow-up in a rabbit model. Journal of Materials Science: Materials in Medicine 31 (12).Date of publication of this fulltext: 11 Oct 2021 12:35
Article
DOI to cite this document: 10.5283/epub.49490
Abstract
This study aimed to investigate in vivo two stent technologies, with particular emphasis on thrombogenicity and inflammatory vessel remodeling processes. The micro-stents tested in this study were developed for intracranial aneurysm treatment. In our study twelve, New Zealand white rabbits were divided into two groups: 18 laser-cut stents (LCS) and 18 braided stents (BS) were impanated without ...
This study aimed to investigate in vivo two stent technologies, with particular emphasis on thrombogenicity and inflammatory vessel remodeling processes. The micro-stents tested in this study were developed for intracranial aneurysm treatment. In our study twelve, New Zealand white rabbits were divided into two groups: 18 laser-cut stents (LCS) and 18 braided stents (BS) were impanated without admiration of antiplatelet medication. Three stents were implanted into each animal in the common carotid artery, subclavian artery, and abdominal aorta. Digital subtraction angiography was performed before and after stent implantation and at follow-up for the visualization of occurring In-stent thromboembolism or stenosis. The Stents were explanted for histopathological examination at two different timepoints, after 3 and 28 days. Angiographically neither in-stent thrombosis nor stenosis for both groups was seen. There was a progressive increase in the vessel diameter, which was more pronounced for BS than for LCS. We detected a higher number of thrombi adherent to the foreign material on day 3 for BS. On day 3, the neointima was absent, whereas the complete formation observed was on day 28. There was no significant difference between both groups regarding the thickness of the neointima. The in vivo model of our study enabled the evaluation of blood and vessel reactions for two different stent technologies. Differences in vessel dimension and tissue around the stents were observed on day 28. Histological analysis on day 3 enabled the assessment of thrombotic reactions, representing an important complementary result in long-term studies. [GRAPHICS] .
Alternative links to fulltext
Involved Institutions
Details
| Item type | Article | ||||
| Journal or Publication Title | Journal of Materials Science: Materials in Medicine | ||||
| Publisher: | Springer | ||||
|---|---|---|---|---|---|
| Open Access Type: | DEAL (Springer) - Non UR | ||||
| Place of Publication: | DORDRECHT | ||||
| Volume: | 31 | ||||
| Number of Issue or Book Chapter: | 12 | ||||
| Date | 3 December 2020 | ||||
| Institutions | Medicine > Lehrstuhl für Pathologie | ||||
| Identification Number |
| ||||
| Keywords | In vivo comparison of braided and laser-cut stents; Thromboembolism; Neointima | ||||
| Dewey Decimal Classification | 600 Technology > 610 Medical sciences Medicine | ||||
| Status | Published | ||||
| Refereed | Yes, this version has been refereed | ||||
| Created at the University of Regensburg | Yes | ||||
| URN of the UB Regensburg | urn:nbn:de:bvb:355-epub-494909 | ||||
| Item ID | 49490 |