Holzamer, Andreas ; Bedogni, Francesco ; van Wyk, Pieter ; Barwad, Parag ; Protasiewicz, Marcin ; Ielasi, Alfonso ; Nombela‐Franco, Luis ; Seidler, Tim ; Hilker, Michael

Performance of the 32 mm Myval transcatheter heart valve for treatment of aortic stenosis in patients with extremely large aortic annuli in real‐world scenario: First global, multicenter experience

Holzamer, Andreas , Bedogni, Francesco, van Wyk, Pieter, Barwad, Parag, Protasiewicz, Marcin, Ielasi, Alfonso, Nombela‐Franco, Luis, Seidler, Tim und Hilker, Michael (2023) Performance of the 32 mm Myval transcatheter heart valve for treatment of aortic stenosis in patients with extremely large aortic annuli in real‐world scenario: First global, multicenter experience. Catheterization and Cardiovascular Interventions.

Veröffentlichungsdatum dieses Volltextes: 19 Sep 2023 09:25
Artikel
DOI zum Zitieren dieses Dokuments: 10.5283/epub.54699


Zusammenfassung

Background Extremely large aortic valve anatomy is one of the remaining limitations leading to exclusion of patients for transcatheter aortic valve replacement (TAVR). Aims The newly approved Myval 32 mm device is designed for use in aortic annulus areas up to 840 mm2. Here we want to share the initial worldwide experience with the device. Methods and Results Retrospective data were ...

Background
Extremely large aortic valve anatomy is one of the remaining limitations leading to exclusion of patients for transcatheter aortic valve replacement (TAVR).
Aims
The newly approved Myval 32 mm device is designed for use in aortic annulus areas up to 840 mm2. Here we want to share the initial worldwide experience with the device.
Methods and Results
Retrospective data were collected from 10 patients with aortic stenosis and very large annular anatomy (mean area 765.5 mm2), who underwent implantation with 32 mm Myval transcatheter heart valve at eight centers. Valve Academic Research Consortium-2 device success was achieved in all cases. Mild paravalvular leak was observed in three patients and two patients required new pacemaker implantation. One patient experienced retroperitoneal hemorrhage caused by the contralateral 6 F sheath and required surgical revision. No device-related complications, stroke, or death from any cause occurred within the 30-day follow-up period. In a studied cohort of 2219 consecutive TAVR-screened patients from a central European site, only 0.27% of patients showed larger anatomy than covered by the 32 mm Myval device by instructions for use without off-label use of overexpansion. This rate was significantly higher for the 34 mm Evolut Pro (1.8%) and 29 mm Sapien 3 (2.1%) devices.
Conclusions
The Myval 32 mm prosthesis showed promising initial results in a cohort of patients who previously had to be excluded from TAVR. It is desirable that all future TAVR systems accommodate larger anatomy to allow optimal treatment of all patients.



Beteiligte Einrichtungen


Details

DokumentenartArtikel
Titel eines Journals oder einer ZeitschriftCatheterization and Cardiovascular Interventions
Verlag:Wiley
Datum12 September 2023
InstitutionenMedizin > Lehrstuhl für Herz-, Thorax- und herznahe Gefäßchirurgie
Identifikationsnummer
WertTyp
10.1002/ccd.30820DOI
Stichwörter / Keywordsaortic valve disease, new devices (in general), percutaneous valve therapy
Dewey-Dezimal-Klassifikation600 Technik, Medizin, angewandte Wissenschaften > 610 Medizin
StatusVeröffentlicht
BegutachtetJa, diese Version wurde begutachtet
An der Universität Regensburg entstandenZum Teil
URN der UB Regensburgurn:nbn:de:bvb:355-epub-546994
Dokumenten-ID54699

Bibliographische Daten exportieren

Nur für Besitzer und Autoren: Kontrollseite des Eintrags

nach oben