Dobler, Barbara ; Mai, Sabine ; Ross, Christine ; Wolff, Dirk ; Wertz, Hansjörg ; Lohr, Frank ; Wenz, Frederik

Evaluation of possible prostate displacement induced by pressure applied during transabdominal ultrasound image acquisition

Dobler, Barbara, Mai, Sabine, Ross, Christine, Wolff, Dirk, Wertz, Hansjörg, Lohr, Frank und Wenz, Frederik (2006) Evaluation of possible prostate displacement induced by pressure applied during transabdominal ultrasound image acquisition. Strahlentherapie und Onkologie : Organ der Deutschen Röntgengesellschaft ... [et al] 182 (4), S. 240-246.

Veröffentlichungsdatum dieses Volltextes: 05 Aug 2009 13:26
Artikel
DOI zum Zitieren dieses Dokuments: 10.5283/epub.1221


Zusammenfassung

Background and Purpose: For accurate positioning of the prostate in external radiotherapy, transabdominal ultrasound Localization and positioning systems are available. Reports have stated that probe pressure applied during image acquisition causes clinically relevant prostate displacement. The aim of this study was to investigate the prostate displacement due to the pressure applied during ...

Background and Purpose: For accurate positioning of the prostate in external radiotherapy, transabdominal ultrasound Localization and positioning systems are available. Reports have stated that probe pressure applied during image acquisition causes clinically relevant prostate displacement. The aim of this study was to investigate the prostate displacement due to the pressure applied during transabdominal ultrasound image acquisition with the BAT((R)) ultrasound system. Material and Methods: For ten patients who had undergone iodine-125 seed implantation for brachytherapy of prostate cancer, X-ray simulations were performed before and during ultrasound image acquisition. The iodine seeds are visible on the X-ray images, representing the position of the prostate. The simulator's crosshair, indicating the isocenter, was used as reference coordinate system. For each patient the change in prostate position was calculated based on the seed positions during and after ultrasound examination. Results: A maximum displacement of the prostate of 2.3 mm in anteroposterior and 1.9 mm in craniocaudal direction and a rotational change of up to 2.5 degrees were observed. If the system was not handled correctly and too much pressure was applied, a shift of the prostate of up to 10 mm could be induced. Conclusion: Compared to the prostate displacement due to changes in rectal filling, which according to Crook et al. can be as much as 1.7 cm, the maximum displacement of Less than 0.3 cm caused by the probe pressure is negligible. However, proper education of the staff and preparation of the patient are essential for the safe use of the system.



Beteiligte Einrichtungen


Details

DokumentenartArtikel
Titel eines Journals oder einer ZeitschriftStrahlentherapie und Onkologie : Organ der Deutschen Röntgengesellschaft ... [et al]
Verlag:URBAN & VOGEL
Ort der Veröffentlichung:MUNICH
Band:182
Nummer des Zeitschriftenheftes oder des Kapitels:4
Seitenbereich:S. 240-246
DatumApril 2006
InstitutionenMedizin > Lehrstuhl für Strahlentherapie
Identifikationsnummer
WertTyp
16622626PubMed-ID
10.1007/s00066-006-1513-0DOI
Stichwörter / KeywordsINTENSITY-MODULATED RADIOTHERAPY; EXTERNAL-BEAM RADIOTHERAPY; RADIATION-THERAPY; CANCER; LOCALIZATION; PATIENT; IMRT; MOTION; ERROR; BRACHYTHERAPY; ultrasound localization; image-guided radiotherapy; prostate motion; ultrasound probe pressure; patient positioning
Dewey-Dezimal-Klassifikation600 Technik, Medizin, angewandte Wissenschaften > 610 Medizin
StatusVeröffentlicht
BegutachtetJa, diese Version wurde begutachtet
An der Universität Regensburg entstandenUnbekannt / Keine Angabe
Dokumenten-ID1221

Bibliographische Daten exportieren

Nur für Besitzer und Autoren: Kontrollseite des Eintrags

nach oben