| Download ( PDF | 153kB) Repository staff only |
Characterization of a reservoir-type capillary optical microsensor for pCO2 measurements
Ertekin, Kadriye, Klimant, Ingo, Neurauter, Gerhard and Wolfbeis, Otto S.
(2003)
Characterization of a reservoir-type capillary optical microsensor for pCO2 measurements.
Talanta 59 (2), pp. 261-263.
Date of publication of this fulltext: 20 May 2011 06:20
Article
DOI to cite this document: 10.5283/epub.20834
Abstract
A reservoir type of capillary microsensor for pCO(2) measurements is presented. The sensor is based on the measurement of the fluorescence intensity of the anionic form of the pH indicator 1-hydroxy-3,6,8-pyrenetrisulfonate in the form of its ion pair with a quaternary ammonium base in an ethyl cellulose matrix. The glass capillary containing the reservoir sensor was prepared by immersing the tip ...
A reservoir type of capillary microsensor for pCO(2) measurements is presented. The sensor is based on the measurement of the fluorescence intensity of the anionic form of the pH indicator 1-hydroxy-3,6,8-pyrenetrisulfonate in the form of its ion pair with a quaternary ammonium base in an ethyl cellulose matrix. The glass capillary containing the reservoir sensor was prepared by immersing the tip of the optical fiber into the sensing agent very close to the sensor tip thus providing a very small volume for the sensing reaction. The purpose of the sensing approach is to regenerate the dye/buffer system by diffusion, which may be poisoned by interfering acids, or bleach by photolysis. The fresh cocktail from the reservoir takes the place of protonated form of the dye. The internal buffer system also makes the protonation-deprotonation equilibria reversible. The distal tip of the internal buffer containing reservoir is coated with a gas-permeable but ion-impermeable teflon membrane. The dynamic range for the detection of PCO2 is between 1 and 20 hPa, which corresponds to the range of dissolved CO2 in water. The response time is 15 s and the detection limit is 1 hPa of PCO2. The recovery performance of this sensor can be improved by means of mechanical adjustment of the sensor tip in a micrometric scale. (C) 2002 Elsevier Science B.V. All rights reserved.
Alternative links to fulltext
Involved Institutions
Details
| Item type | Article | ||||
| Journal or Publication Title | Talanta | ||||
| Publisher: | ELSEVIER SCIENCE BV | ||||
|---|---|---|---|---|---|
| Place of Publication: | AMSTERDAM | ||||
| Volume: | 59 | ||||
| Number of Issue or Book Chapter: | 2 | ||||
| Page Range: | pp. 261-263 | ||||
| Date | 2003 | ||||
| Institutions | Chemistry and Pharmacy > Institut für Analytische Chemie, Chemo- und Biosensorik | ||||
| Identification Number |
| ||||
| Keywords | CARBON-DIOXIDE SENSOR; FLUORESCENCE; OXYGEN; CO2; PH; carbon dioxide; fiber optic chemical sensor; capillary microsensor | ||||
| Dewey Decimal Classification | 500 Science > 540 Chemistry & allied sciences | ||||
| Status | Published | ||||
| Refereed | Unknown | ||||
| Created at the University of Regensburg | Unknown | ||||
| Item ID | 20834 |
Download Statistics
Download Statistics