Commissioning of volumetric modulated arc therapy (VMAT) in a dual-vendor environment

Dobler, Barbara, Groeger, Christian, Treutwein, Marius, Alvarez Moret, Judith, Goetzfried, Thomas, Weidner, Karin, Härtl, Petra und Koelbl, Oliver (2011) Commissioning of volumetric modulated arc therapy (VMAT) in a dual-vendor environment. Radiotherapy and Oncology (99), S. 86-89.

[img]
Vorschau
PDF - Angenommene Version
Volltext
Download (228kB)

Zum Artikel beim Verlag (über DOI)

Zum PubMed-Eintrag dieses Artikels

Andere URL zum Volltext: http://dx.doi.org/10.1016/j.radonc.2011.01.024


Zusammenfassung

Methods and results for commissioning of the complete VMAT delivery chain are presented for the combination of Nucletron's Oncentra MasterPlan® v3.3 with Elekta's Mosaiq® v1.6 and SynergyS® linac. VMAT specific linac commissioning included determination of the size of the minimal dynamic leaf gap. Dosimetric validation of the complete treatment chain was performed using a 2D-ionization-chamber-array and showed excellent dosimetric results.


Bibliographische Daten exportieren



Dokumentenart:Artikel
Datum:2011
Institutionen:Medizin > Lehrstuhl für Strahlentherapie
Identifikationsnummer:
WertTyp
10.1016/j.radonc.2011.01.024DOI
21458090PubMed-ID
Stichwörter / Keywords:Volumetric modulated arc therapy (VMAT); Commissioning; Clinical implementation
Dewey-Dezimal-Klassifikation:600 Technik, Medizin, angewandte Wissenschaften > 610 Medizin
Status:Veröffentlicht
Begutachtet:Ja, diese Version wurde begutachtet
An der Universität Regensburg entstanden:Ja
Eingebracht am:06 Feb 2012 13:28
Zuletzt geändert:19 Dez 2016 19:55
Dokumenten-ID:23390
Nur für Besitzer und Autoren: Kontrollseite des Eintrags
  1. Universität

Universitätsbibliothek

Publikationsserver

Kontakt:

Publizieren: oa@ur.de

Dissertationen: dissertationen@ur.de

Forschungsdaten: daten@ur.de

Ansprechpartner