Commissioning of volumetric modulated arc therapy (VMAT) in a dual-vendor environment

Dobler, Barbara, Groeger, Christian, Treutwein, Marius, Alvarez Moret, Judith, Goetzfried, Thomas, Weidner, Karin, Härtl, Petra and Koelbl, Oliver (2011) Commissioning of volumetric modulated arc therapy (VMAT) in a dual-vendor environment. Radiotherapy and Oncology (99), pp. 86-89.

[img]
Preview
PDF - Accepted Version
Volltext
Download (228kB)
Date of publication of this fulltext: 06 Feb 2012 13:28

at publisher (via DOI)

at PubMed

Other URL: http://dx.doi.org/10.1016/j.radonc.2011.01.024


Abstract

Methods and results for commissioning of the complete VMAT delivery chain are presented for the combination of Nucletron's Oncentra MasterPlan® v3.3 with Elekta's Mosaiq® v1.6 and SynergyS® linac. VMAT specific linac commissioning included determination of the size of the minimal dynamic leaf gap. Dosimetric validation of the complete treatment chain was performed using a 2D-ionization-chamber-array and showed excellent dosimetric results.


Export bibliographical data



Item type:Article
Date:2011
Institutions:Medicine > Lehrstuhl für Strahlentherapie
Identification Number:
ValueType
10.1016/j.radonc.2011.01.024DOI
21458090PubMed ID
Keywords:Volumetric modulated arc therapy (VMAT); Commissioning; Clinical implementation
Dewey Decimal Classification:600 Technology > 610 Medical sciences Medicine
Status:Published
Refereed:Yes, this version has been refereed
Created at the University of Regensburg:Yes
Item ID:23390
Owner only: item control page

Downloads

Downloads per month over past year

  1. Homepage UR

University Library

Publication Server

Contact:

Publishing: oa@ur.de

Dissertations: dissertationen@ur.de

Research data: daten@ur.de

Contact persons