Prasser, Christopher ; Bele, Sylvia ; Keyl, Cornelius ; Schweiger, Stefan ; Trabold, Benedikt ; Amann, Matthias ; Welnhofer, Julia ; Wiesenack, Christoph

Evaluation of a new arterial pressure-based cardiac output device requiring no external calibration

Prasser, Christopher, Bele, Sylvia , Keyl, Cornelius, Schweiger, Stefan, Trabold, Benedikt, Amann, Matthias, Welnhofer, Julia and Wiesenack, Christoph (2007) Evaluation of a new arterial pressure-based cardiac output device requiring no external calibration. BMC Anesthesiology 7 (9).

Date of publication of this fulltext: 09 Feb 2017 14:30
Article
DOI to cite this document: 10.5283/epub.35182


Abstract

Background Several techniques have been discussed as alternatives to the intermittent bolus thermodilution cardiac output (COPAC) measurement by the pulmonary artery catheter (PAC). However, these techniques usually require a central venous line, an additional catheter, or a special calibration procedure. A new arterial pressure-based cardiac output (COAP) device (FloTrac™, Vigileo™; Edwards ...

Background
Several techniques have been discussed as alternatives to the intermittent bolus thermodilution cardiac output (COPAC) measurement by the pulmonary artery catheter (PAC). However, these techniques usually require a central venous line, an additional catheter, or a special calibration procedure. A new arterial pressure-based cardiac output (COAP) device (FloTrac™, Vigileo™; Edwards Lifesciences, Irvine, CA, USA) only requires access to the radial or femoral artery using a standard arterial catheter and does not need an external calibration. We validated this technique in critically ill patients in the intensive care unit (ICU) using COPAC as the method of reference.
Methods
We studied 20 critically ill patients, aged 16 to 74 years (mean, 55.5 ± 18.8 years), who required both arterial and pulmonary artery pressure monitoring. COPAC measurements were performed at least every 4 hours and calculated as the average of 3 measurements, while COAP values were taken immediately at the end of bolus determinations. Accuracy of measurements was assessed by calculating the bias and limits of agreement using the method described by Bland and Altman.
Results
A total of 164 coupled measurements were obtained. Absolute values of COPAC ranged from 2.80 to 10.80 l/min (mean 5.93 ± 1.55 l/min). The bias and limits of agreement between COPAC and COAP for unequal numbers of replicates was 0.02 ± 2.92 l/min. The percentage error between COPAC and COAP was 49.3%. The bias between percentage changes in COPAC (ΔCOPAC) and percentage changes in COAP (ΔCOAP) for consecutive measurements was -0.70% ± 32.28%. COPAC and COAP showed a Pearson correlation coefficient of 0.58 (p < 0.01), while the correlation coefficient between ΔCOPAC and ΔCOAP was 0.46 (p < 0.01).
Conclusion
Although the COAP algorithm shows a minimal bias with COPAC over a wide range of values in an inhomogeneous group of critically ill patients, the scattering of the data remains relative wide. Therefore, the used algorithm (V 1.03) failed to demonstrate an acceptable accuracy in comparison to the clinical standard of cardiac output determination.



Involved Institutions


Details

Item typeArticle
Journal or Publication TitleBMC Anesthesiology
Publisher:BMC
Open Access Type:Gold (mit APC - nicht UR)
Volume:7
Number of Issue or Book Chapter:9
Date9 November 2007
InstitutionsMedicine > Lehrstuhl für Anästhesiologie
Identification Number
ValueType
10.1186/1471-2253-7-9DOI
Dewey Decimal Classification600 Technology > 610 Medical sciences Medicine
StatusPublished
RefereedYes, this version has been refereed
Created at the University of RegensburgYes
URN of the UB Regensburgurn:nbn:de:bvb:355-epub-351822
Item ID35182

Export bibliographical data

Owner only: item control page

nach oben