Evaluation of a new arterial pressure-based cardiac output device requiring no external calibration
Prasser, Christopher, Bele, Sylvia
, Keyl, Cornelius, Schweiger, Stefan, Trabold, Benedikt, Amann, Matthias, Welnhofer, Julia und Wiesenack, Christoph
(2007)
Evaluation of a new arterial pressure-based cardiac output device requiring no external calibration.
BMC Anesthesiology 7 (9).
Veröffentlichungsdatum dieses Volltextes: 09 Feb 2017 14:30
Artikel
DOI zum Zitieren dieses Dokuments: 10.5283/epub.35182
Zusammenfassung
Background Several techniques have been discussed as alternatives to the intermittent bolus thermodilution cardiac output (COPAC) measurement by the pulmonary artery catheter (PAC). However, these techniques usually require a central venous line, an additional catheter, or a special calibration procedure. A new arterial pressure-based cardiac output (COAP) device (FloTrac™, Vigileo™; Edwards ...
Background
Several techniques have been discussed as alternatives to the intermittent bolus thermodilution cardiac output (COPAC) measurement by the pulmonary artery catheter (PAC). However, these techniques usually require a central venous line, an additional catheter, or a special calibration procedure. A new arterial pressure-based cardiac output (COAP) device (FloTrac™, Vigileo™; Edwards Lifesciences, Irvine, CA, USA) only requires access to the radial or femoral artery using a standard arterial catheter and does not need an external calibration. We validated this technique in critically ill patients in the intensive care unit (ICU) using COPAC as the method of reference.
Methods
We studied 20 critically ill patients, aged 16 to 74 years (mean, 55.5 ± 18.8 years), who required both arterial and pulmonary artery pressure monitoring. COPAC measurements were performed at least every 4 hours and calculated as the average of 3 measurements, while COAP values were taken immediately at the end of bolus determinations. Accuracy of measurements was assessed by calculating the bias and limits of agreement using the method described by Bland and Altman.
Results
A total of 164 coupled measurements were obtained. Absolute values of COPAC ranged from 2.80 to 10.80 l/min (mean 5.93 ± 1.55 l/min). The bias and limits of agreement between COPAC and COAP for unequal numbers of replicates was 0.02 ± 2.92 l/min. The percentage error between COPAC and COAP was 49.3%. The bias between percentage changes in COPAC (ΔCOPAC) and percentage changes in COAP (ΔCOAP) for consecutive measurements was -0.70% ± 32.28%. COPAC and COAP showed a Pearson correlation coefficient of 0.58 (p < 0.01), while the correlation coefficient between ΔCOPAC and ΔCOAP was 0.46 (p < 0.01).
Conclusion
Although the COAP algorithm shows a minimal bias with COPAC over a wide range of values in an inhomogeneous group of critically ill patients, the scattering of the data remains relative wide. Therefore, the used algorithm (V 1.03) failed to demonstrate an acceptable accuracy in comparison to the clinical standard of cardiac output determination.
Alternative Links zum Volltext
Beteiligte Einrichtungen
Details
| Dokumentenart | Artikel | ||||
| Titel eines Journals oder einer Zeitschrift | BMC Anesthesiology | ||||
| Verlag: | BMC | ||||
|---|---|---|---|---|---|
| Band: | 7 | ||||
| Nummer des Zeitschriftenheftes oder des Kapitels: | 9 | ||||
| Datum | 9 November 2007 | ||||
| Institutionen | Medizin > Lehrstuhl für Anästhesiologie | ||||
| Identifikationsnummer |
| ||||
| Dewey-Dezimal-Klassifikation | 600 Technik, Medizin, angewandte Wissenschaften > 610 Medizin | ||||
| Status | Veröffentlicht | ||||
| Begutachtet | Ja, diese Version wurde begutachtet | ||||
| An der Universität Regensburg entstanden | Ja | ||||
| URN der UB Regensburg | urn:nbn:de:bvb:355-epub-351822 | ||||
| Dokumenten-ID | 35182 |
Downloadstatistik
Downloadstatistik