Direkt zum Inhalt

Treutwein, Marius ; Hipp, Matthias ; Koelbl, Oliver ; Dobler, Barbara

Volumetric-modulated arc therapy and intensity-modulated radiation therapy treatment planning for prostate cancer with flattened beam and flattening filter free linear accelerators

Treutwein, Marius , Hipp, Matthias, Koelbl, Oliver und Dobler, Barbara (2017) Volumetric-modulated arc therapy and intensity-modulated radiation therapy treatment planning for prostate cancer with flattened beam and flattening filter free linear accelerators. Journal of Applied Clinical Medical Physics 18 (5), S. 307-314.

Veröffentlichungsdatum dieses Volltextes: 04 Sep 2017 05:44
Artikel
DOI zum Zitieren dieses Dokuments: 10.5283/epub.36127


Zusammenfassung

This study on patients with localized prostate cancer was set up to investigate valuable differences using flattened beam (FB) and flattening filter free (FFF) mode in the application of intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT). For ten patients, four different plans were calculated with Oncentra planning system of Elekta, using Synergy machines: IMRT ...

This study on patients with localized prostate cancer was set up to investigate valuable differences using flattened beam (FB) and flattening filter free (FFF) mode in the application of intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT). For ten patients, four different plans were calculated with Oncentra planning system of Elekta, using Synergy machines: IMRT and VMAT, with and without flattening filter. Homogeneity and conformity indexes, dose to the organs at risk, and measurements of peripheral dose and dosimetric plan verification including record of the delivery times were analyzed and statistically evaluated. The indexes for homogeneity and conformity (CTV and PTV) are either advantageous or not significantly different for FFF compared to FB with one minor exception. Regarding the doses to the organs at risk and the measured peripheral dose, equivalent or lower doses were delivered for FFF than with FB. Furthermore, the delivery times were significantly shorter for FFF. VMAT compared to IMRT reveals benefits or at least equivalent values. VMAT-FFF combines the most advantageous plan quality parameters with the shortest delivery times and reduced peripheral dose and is therefore recommended for the given equipment and cancer localization.



Beteiligte Einrichtungen


Details

DokumentenartArtikel
Titel eines Journals oder einer ZeitschriftJournal of Applied Clinical Medical Physics
Verlag:Wiley
Ort der Veröffentlichung:HOBOKEN
Band:18
Nummer des Zeitschriftenheftes oder des Kapitels:5
Seitenbereich:S. 307-314
Datum30 August 2017
InstitutionenMedizin > Lehrstuhl für Strahlentherapie
Identifikationsnummer
WertTyp
10.1002/acm2.12168DOI
28857432PubMed-ID
Stichwörter / Keywords3-DIMENSIONAL CONFORMAL RADIOTHERAPY; SIMULTANEOUS INTEGRATED BOOST; INTRAFRACTION MOTION; TREATMENT PLANS; IMRT; VMAT; OPTIMIZATION; flattening filter free; IMRT; prostate cancer; VMAT
Dewey-Dezimal-Klassifikation600 Technik, Medizin, angewandte Wissenschaften > 610 Medizin
600 Technik, Medizin, angewandte Wissenschaften > 610 Medizin
StatusVeröffentlicht