Direkt zum Inhalt

Pregler, Benedikt ; Beyer, L. P. ; Teufel, A. ; Niessen, C. ; Stroszczynski, C. ; Brodoefel, H. ; Wiggermann, P.

Low Tube Voltage Liver MDCT with Sinogram-Affirmed Iterative Reconstructions for the Detection of Hepatocellular Carcinoma

Article

Pregler, Benedikt, Beyer, L. P., Teufel, A., Niessen, C., Stroszczynski, C., Brodoefel, H. and Wiggermann, P. (2017) Low Tube Voltage Liver MDCT with Sinogram-Affirmed Iterative Reconstructions for the Detection of Hepatocellular Carcinoma. Scientific Reports 7 (1), p. 9460.

DOI to cite this document: 10.5283/epub.36533


Abstract

Aim of this study was to compare low tube voltage computed tomography (80 kV) of the liver using iterative image reconstruction (SAFIRE) with standard computed tomography (120 kV) using filtered back-projection (FBP) for the detection of hepatocellular carcinoma (HCC). 46 patients (43 men) with 93 HCC confirmed by 3T MRI with Gd-EOB-DPTA, in inconclusive cases combined with contrast-enhanced ...

Aim of this study was to compare low tube voltage computed tomography (80 kV) of the liver using iterative image reconstruction (SAFIRE) with standard computed tomography (120 kV) using filtered back-projection (FBP) for the detection of hepatocellular carcinoma (HCC). 46 patients (43 men) with 93 HCC confirmed by 3T MRI with Gd-EOB-DPTA, in inconclusive cases combined with contrast-enhanced ultrasound, underwent dual-energy CT. The raw data of the 80 kV tube was reconstructed using the iterative reconstruction algorithm SAFIRE with two strengths (I3 and I5). The virtual 120 kV image data set was reconstructed using FBP. The CT images were reviewed to determine the lesion-to-liver contrast (LLC), the lesion contrast-to-noise ratio (CNR) and the sensitivity. The LLC (57.1/54.3 [I3/I5] vs. 34.9 [FBP]; p <= 0.01), CNR (3.67/4.45 [I3/I5] vs. 2.48 [FBP]; p < 0.01) and sensitivity (91.4%/88.2% [I3/I5] vs. 72.0% [FBP]; p <= 0.01) were significantly higher in the low-voltage protocol using SAFIRE. Therefore, low tube voltage CT using SAFIRE results in an increased lesion-to-liver contrast as well as an improved lesion contrast-to-noise ratio compared to FBP at 120 kV which results in a higher sensitivity for the detection of HCC.



Involved Institutions


Details

Item typeArticle
Journal or Publication TitleScientific Reports
PublisherNature
Open Access TypeGold (with APC)
Place of PublicationLONDON
Volume7
Number of Issue or Book Chapter1
Page Rangep. 9460
Date25 August 2017
Date of publication17 Jan 2018 14:53
InstitutionsMedicine > Lehrstuhl für Innere Medizin I
Medicine > Lehrstuhl für Röntgendiagnostik
Identification Number
ValueType
10.1038/s41598-017-10095-6DOI
KeywordsABDOMINAL COMPUTED-TOMOGRAPHY; FILTERED BACK-PROJECTION; RADIATION-DOSE REDUCTION; IMAGE QUALITY; MULTIDETECTOR CT; PHANTOM; ALGORITHM; SAFIRE; NOISE; ANGIOGRAPHY;
Dewey Decimal Classification600 Technology > 610 Medical sciences Medicine
StatusPublished
RefereedYes, this version has been refereed
Created at the University of RegensburgYes
URN of the UB Regensburgurn:nbn:de:bvb:355-epub-365330
Item ID36533

Export bibliographical data

Owner only: item control page

nach oben