Direkt zum Inhalt

Bauer, Meike ; Wunderlich, Lukas ; Weinzierl, Florian ; Duerkop, Axel ; Baeumner, Antje J. ; ;

Electrochemical multi-analyte point-of-care perspiration sensors using on-chip three-dimensional graphene electrodes

Bauer, Meike, Wunderlich, Lukas, Weinzierl, Florian, Duerkop, Axel , Baeumner, Antje J., make_name_string expected hash reference and make_name_string expected hash reference (2021) Electrochemical multi-analyte point-of-care perspiration sensors using on-chip three-dimensional graphene electrodes. Analytical and Bioanalytical Chemistry 413, pp. 763-777.

Date of publication of this fulltext: 05 Feb 2021 06:27
Article
DOI to cite this document: 10.5283/epub.44761


Abstract

Multi-analyte sensing using exclusively laser-induced graphene (LIG)-based planar electrode systems was developed for sweat analysis. LIG provides 3D structures of graphene, can be manufactured easier than any other carbon electrode also on large scale, and in form of electrodes: hence, it is predestinated for affordable, wearable point-of-care sensors. Here, it is demonstrated that LIG ...

Multi-analyte sensing using exclusively laser-induced graphene (LIG)-based planar electrode systems was developed for sweat analysis. LIG provides 3D structures of graphene, can be manufactured easier than any other carbon electrode also on large scale, and in form of electrodes: hence, it is predestinated for affordable, wearable point-of-care sensors. Here, it is demonstrated that LIG facilitates all three electrochemical sensing strategies (voltammetry, potentiometry, impedance) in a multi-analyte system for sweat analysis. A potentiometric potassium-ion-selective electrode in combination with an electrodeposited Ag/AgCl reference electrode (RE) enabled the detection of potassium ions in the entire physiologically relevant range (1 to 500 mM) with a fast response time, unaffected by the presence of main interfering ions and sweat-collecting materials. A kidney-shaped interdigitated LIG electrode enabled the determination of the overall electrolyte concentration by electrochemical impedance spectroscopy at a fixed frequency. Enzyme-based strategies with amperometric detection share a common RE and were realized with Prussian blue as electron mediator and biocompatible chitosan for enzyme immobilization and protection of the electrode. Using glucose and lactate oxidases, lower limits of detection of 13.7 +/- 0.5 mu M for glucose and 28 +/- 3 mu M for lactate were obtained, respectively. The sensor showed a good performance at different pH, with sweat-collecting tissues, on a model skin system and furthermore in synthetic sweat as well as in artificial tear fluid. Response time for each analytical cycle totals 75 s, and hence allows a quasi-continuous and simultaneous monitoring of all analytes. This multi-analyte all-LIG system is therefore a practical, versatile, and most simple strategy for point-of-care applications and has the potential to outcompete standard screen-printed electrodes.



Involved Institutions


Details

Item typeArticle
Journal or Publication TitleAnalytical and Bioanalytical Chemistry
Publisher:SPRINGER HEIDELBERG
Place of Publication:HEIDELBERG
Volume:413
Page Range:pp. 763-777
Date2021
InstitutionsChemistry and Pharmacy > Institut für Analytische Chemie, Chemo- und Biosensorik
Chemistry and Pharmacy > Institut für Analytische Chemie, Chemo- und Biosensorik > Chemo- und Biosensorik (Prof. Antje J. Bäumner, formerly Prof. Wolfbeis)
Identification Number
ValueType
10.1007/s00216-020-02939-4DOI
KeywordsHUMAN SWEAT; GLUCOSE; BIOSENSORS; LACTATE; FABRICATION; MEMBRANES; Laser-induced graphene (LIG); Point-of-care (POC); Electrochemical biosensor; Sweat sensor; Health-monitoring platform
Dewey Decimal Classification500 Science > 540 Chemistry & allied sciences
StatusPublished
RefereedYes, this version has been refereed
Created at the University of RegensburgYes
URN of the UB Regensburgurn:nbn:de:bvb:355-epub-447618
Item ID44761

Export bibliographical data

Owner only: item control page

nach oben