Mühl-Benninghaus, Ruben ; Fries, Frederik ; Kießling, Mara ; Tomori, Toshiki ; Krajewski, Stefanie ; Simgen, Andreas ; Bauer, Sabina ; Hey, Natascha ; Brynda, Eduard ; Taborska, Johanka ; Riedel, Tomáš ; Reith, Wolfgang ; Cattaneo, Giorgio ; Brochhausen, Christoph

Vascular Response on a Novel Fibrin-Based Coated Flow Diverter

Mühl-Benninghaus, Ruben , Fries, Frederik, Kießling, Mara, Tomori, Toshiki, Krajewski, Stefanie, Simgen, Andreas, Bauer, Sabina, Hey, Natascha, Brynda, Eduard, Taborska, Johanka, Riedel, Tomáš, Reith, Wolfgang, Cattaneo, Giorgio und Brochhausen, Christoph (2021) Vascular Response on a Novel Fibrin-Based Coated Flow Diverter. CardioVascular and Interventional Radiology 45 (2), S. 236-243.

Veröffentlichungsdatum dieses Volltextes: 29 Feb 2024 12:41
Artikel
DOI zum Zitieren dieses Dokuments: 10.5283/epub.56872


Zusammenfassung

Purpose Due to thromboembolic complications and in-stent-stenosis after flow diverter (FD) treatment, the long-term use of dual antiplatelet treatment (DAPT) is mandatory. The tested nano-coating has been shown to reduce material thrombogenicity and promote endothelial cell proliferation in vitro. We compared the biocompatibility of coated (Derivo Heal) and non-coated (Derivo bare) FDs with DAPT ...

Purpose
Due to thromboembolic complications and in-stent-stenosis after flow diverter (FD) treatment, the long-term use of dual antiplatelet treatment (DAPT) is mandatory. The tested nano-coating has been shown to reduce material thrombogenicity and promote endothelial cell proliferation in vitro. We compared the biocompatibility of coated (Derivo Heal) and non-coated (Derivo bare) FDs with DAPT in an animal model.

Methods
Derivo® bare (n = 10) and Derivo® Heal (n = 10) FD were implanted in the common carotid arteries (CCAs) of New Zealand white rabbits. One additional FD, alternately a Derivo bare (n = 5) or Derivo Heal (n = 5), was implanted in the abdominal aorta (AA) for assessment of the patency of branch arteries. Histopathological examinations were performed after 28 days. Angiography was performed before and after FD implantation and at follow-up.

Results
Statistical analysis of the included specimens showed complete endothelialization of all FDs with no significant differences in neointima thickness between Derivo® bare and Derivo® Heal (CCA: p = 0.91; AA: p = 0.59). A significantly reduced number of macrophages in the vessel wall of the Derivo Heal was observed for the CCA (p = 0.02), and significantly reduced fibrin and platelet deposition on the surface of the Derivo Heal was observed for the AA. All branch arteries of the stented aorta remained patent.

Conclusion
In this animal model, the novel fibrin-based coated FD showed a similar blood and tissue compatibility as the non-coated FD.



Beteiligte Einrichtungen


Details

DokumentenartArtikel
Titel eines Journals oder einer ZeitschriftCardioVascular and Interventional Radiology
Verlag:Springer
Ort der Veröffentlichung:NEW YORK
Band:45
Nummer des Zeitschriftenheftes oder des Kapitels:2
Seitenbereich:S. 236-243
Datum16 Dezember 2021
InstitutionenMedizin > Lehrstuhl für Pathologie
Identifikationsnummer
WertTyp
10.1007/s00270-021-03007-9DOI
Stichwörter / KeywordsSTENT-ASSISTED COILING; INTRACRANIAL ANEURYSMS; DIFFERENTIATION; THROMBOGENICITY; DIVERSION; Flow diverter; Coating; Surface modification; Biocompatibility
Dewey-Dezimal-Klassifikation600 Technik, Medizin, angewandte Wissenschaften > 610 Medizin
StatusVeröffentlicht
BegutachtetJa, diese Version wurde begutachtet
An der Universität Regensburg entstandenJa
URN der UB Regensburgurn:nbn:de:bvb:355-epub-568724