Polyethylene liner dissociation in total hip arthroplasty: a retrospective case–control study on a single implant design

URN to cite this document:
urn:nbn:de:bvb:355-epub-589513
DOI to cite this document:
10.5283/epub.58951
Pagano, S. ; Plate, J. F. ; Kappenschneider, T. ; Reinhard, J. ; Scharf, M. ; Maderbacher, G.
[img]License: Creative Commons Attribution 4.0
PDF - Published Version
(1MB)
Date of publication of this fulltext: 20 Aug 2024 09:12



Abstract

Background Modular acetabular components for total hip arthroplasty (THA) provide intraoperative flexibility; however, polyethylene liner dissociation may occur. This study aimed to examine the incidence and causes of liner dissociation associated with a specific acetabular component design at a single centre. Materials and methods A retrospective analysis of 7027 patients who underwent ...