Oxygenator thrombosis: worst case after development of an abnormal pressure gradient - incidence and pathway

Wendel, Hans P ; Philipp, Alois ; Weber, Norbert ; Birnbaum, Dietrich E ; Ziemer, Gerhard



Abstract

The development of an abnormally high pressure gradient (APG) before the membrane oxygenator (MO) is a complication that occurs during some extracorporeal circulation (ECC) procedures. The present study deals with the incidence of an APG and discusses a probable causative pathway by comparing surface-coated and uncoated oxygenation systems. Five thousand six hundred and seventeen adult ECCs were ...