Go to content
UR Home

Simulated Bone Erosions in a Hand Phantom: Detection with Conventional Screen-Film Technology versus Cesium Iodide-Amorphous Silicon Flat-Panel Detector

Strotzer, Michael ; Völk, Markus ; Wild, Thomas ; von Landenberg, Philipp ; Feuerbach, Stefan



Abstract

PURPOSE: To assess the diagnostic performance of an active-matrix flat-panel x-ray detector for reduced-dose imaging of simulated arthritic lesions. MATERIALS AND METHODS: A digital x-ray detector based on cesium iodide and amorphous silicon technology with a panel size of 43 x 43 cm, matrix of 3,000 x 3,000 pixels, pixel size of 143 mu m, and digital output of 14 bits was used. State-of-the-art ...