Strategies To Design pH Optodes with Luminescence Decay Times in the Microsecond Time Regime
Kosch, Ute, Klimant, Ingo, Werner, Tobias und Wolfbeis, Otto S. (1998) Strategies To Design pH Optodes with Luminescence Decay Times in the Microsecond Time Regime. Analytical Chemistry 70 (18), S. 3892-3897.Veröffentlichungsdatum dieses Volltextes: 18 Apr 2011 13:13
Artikel
DOI zum Zitieren dieses Dokuments: 10.5283/epub.20455
Zusammenfassung
We present the first optical pH sensor with luminescence decay times longer than 1 μs. It is based on radiationless energy transfer from a luminescent ruthenium(II) complex as the donor to a colored pH indicator as the acceptor. The metal−ligand complex ruthenium(II) tris-4,4‘-diphenyl-2,2‘-bipyridyl was selected as the donor, and bromothymol blue and a reactive azo dye were selected as ...
We present the first optical pH sensor with luminescence decay times longer than 1 μs. It is based on radiationless energy transfer from a luminescent ruthenium(II) complex as the donor to a colored pH indicator as the acceptor. The metal−ligand complex ruthenium(II) tris-4,4‘-diphenyl-2,2‘-bipyridyl was selected as the donor, and bromothymol blue and a reactive azo dye were selected as pH-sensitive acceptors. Strategies for co-immobilization of transition metal complexes and pH indicators into a hydrogel matrix were evaluated and discussed. pH transition intervals between 7 and 9 allow the measurement in the physiological range and also under marine conditions. Signal changes of up to 50% over 2.5 pH units are observed, depending on the respective acceptor content in the sensing film. Luminescence lifetime measurements of the sensing films were performed in the frequency domain with a blue LED as the light source. The effect of molecular oxygen (acting both as a quenching and a bleaching agent, thereby limiting the accuracy and the long-term stability of such sensors) was found to be crucial for practical applications. Since leaching of the acceptor also limits the stability of the sensing films, several strategies to reduce this effect were evaluated.
Alternative Links zum Volltext
Beteiligte Einrichtungen
Details
| Dokumentenart | Artikel | ||||
| Titel eines Journals oder einer Zeitschrift | Analytical Chemistry | ||||
| Verlag: | American Chemical Society | ||||
|---|---|---|---|---|---|
| Band: | 70 | ||||
| Nummer des Zeitschriftenheftes oder des Kapitels: | 18 | ||||
| Seitenbereich: | S. 3892-3897 | ||||
| Datum | 1998 | ||||
| Institutionen | Chemie und Pharmazie > Institut für Analytische Chemie, Chemo- und Biosensorik > Chemo- und Biosensorik (Prof. Antje J. Bäumner, ehemals Prof. Wolfbeis) | ||||
| Identifikationsnummer |
| ||||
| Dewey-Dezimal-Klassifikation | 500 Naturwissenschaften und Mathematik > 540 Chemie | ||||
| Status | Veröffentlicht | ||||
| Begutachtet | Unbekannt / Keine Angabe | ||||
| An der Universität Regensburg entstanden | Unbekannt / Keine Angabe | ||||
| Dokumenten-ID | 20455 |
Downloadstatistik
Downloadstatistik