Praktische Dosimetrie und Konstanzprüfung zur Einführung der intraoperativen Bestrahlung mit Intrabeam (Zeiss)
Härtl, Petra, Dobler, Barbara, Kölbl, Oliver and Treutwein, Marius
(2009)
Praktische Dosimetrie und Konstanzprüfung zur Einführung der intraoperativen Bestrahlung mit Intrabeam (Zeiss).
Zeitschrift für Medizinische Physik 19 (4), pp. 288-293.
Date of publication of this fulltext: 09 Feb 2012 13:19
Article
DOI to cite this document: 10.5283/epub.23403
Abstract
The check of dosimetry of the intraoperative radiotherapy system Intrabeam (R) is predefined by the manufacture (Zeiss). The purpose of the study was to develop and implement a method to verify the internal dosimetry of Intrabeamg (R) (Zeiss). Additionally the long-term stability of Intrabeam (R) was checked for dose and isotropy. For dose to water measurements an Unidos (R) was combined with a ...
The check of dosimetry of the intraoperative radiotherapy system Intrabeam (R) is predefined by the manufacture (Zeiss). The purpose of the study was to develop and implement a method to verify the internal dosimetry of Intrabeamg (R) (Zeiss). Additionally the long-term stability of Intrabeam (R) was checked for dose and isotropy. For dose to water measurements an Unidos (R) was combined with a soft jet chamber (TM 23342) which was calibrated in water absorbed dose and as a phantom the type 2962 (PTW Freiburg) was used. RW1 plates were inserted as build up material. The applicators were placed in a bag filled with water to consider the side-scattering. At the surface of the applicator there was a mean difference of 3 percent between the dose to water measurement and the internal dosimetry. The constancy of the dose rate showed a mean deviation of 0.3% at the reference point. The analysis of the dose distribution perpendicular to the applicator axis z (reference z-axis) resulted in a mean deviation of -2.7% (x-direction) and -7, 1% (-x-direction) for the x-axis and, respectively -4. 1% (y-direction) and -5.3% (-y-direction) for the y-axis. The proposed method is suitable to verify the absolute dose of Intrabeam (R). The dose values measured by this method were congruent to the dosimetry of the manu-facture (Zeiss). From out-point of view it is sufficient to verify the absolute dosimetry only at time of commissioning of the system or in the case of changing the applicator. For the daily routine the check of constancy specified by the manufacture is adequate, because the close rate is checked on a daily basis. Additionally the test of constancy showed a high long-term stability in terms of close rate and symmetry.
Involved Institutions
Details
| Item type | Article | ||||||
| Journal or Publication Title | Zeitschrift für Medizinische Physik | ||||||
| Publisher: | ELSEVIER GMBH, URBAN & FISCHER VERLAG | ||||||
|---|---|---|---|---|---|---|---|
| Place of Publication: | JENA | ||||||
| Volume: | 19 | ||||||
| Number of Issue or Book Chapter: | 4 | ||||||
| Page Range: | pp. 288-293 | ||||||
| Date | 2 March 2009 | ||||||
| Additional Information (public) | Notice: this is the author’s version of a work that was accepted for publication in <Zeitschrift für Medizinische Physik>. Changes resulting from the publishing process, such as peer review, editing, corrections, structural formatting, and other quality control mechanisms may not be reflected in this document. Changes may have been made to this work since it was submitted for publication. A definitive version was subsequently published in Zeitschrift für Medizinische Physik, VOL 19, ISSUE 4, 2009, DOI 10.1016/j.zemedi.2009.09.001 | ||||||
| Institutions | Medicine > Lehrstuhl für Strahlentherapie | ||||||
| Identification Number |
| ||||||
| Keywords | STAGE BREAST-CANCER; IORT; Breast cancer; intraoperative Radiotherapy (IORT); Intrabeam | ||||||
| Dewey Decimal Classification | 600 Technology > 610 Medical sciences Medicine 600 Technology > 610 Medical sciences Medicine | ||||||
| Status | Published | ||||||
| Refereed | Yes, this version has been refereed | ||||||
| Created at the University of Regensburg | Yes | ||||||
| Item ID | 23403 |
Download Statistics
Download Statistics