Go to content
UR Home

GC analysis of trichloroacetic acid in water samples by large-volume injection and thermal decarboxylation in a programmed-temperature vaporizer

Drechsel, D., Dettmer, Katja, Engewald, W., Bittner, T. and Efner, J. (2001) GC analysis of trichloroacetic acid in water samples by large-volume injection and thermal decarboxylation in a programmed-temperature vaporizer. Chromatographia 54, pp. 151-154.

Full text not available from this repository.

at publisher (via DOI)


Abstract

AGC method has been developed for the analysis of tricholoroacetic acid (TCA) in water samples. It entails large-volume injection (LVI) and programmed-temperature vaporization (PTV) of water samples, thermal decarboxylation of TCA to chloroform in the injector, and GC-ECD analysis of the decarboxylation product. Conditions such as final injector temperature, splitless time, and injection volume ...

plus


Export bibliographical data



Item type:Article
Date:2001
Institutions:Medicine > Institut für Funktionelle Genomik > Lehrstuhl für Funktionelle Genomik (Prof. Oefner)
Identification Number:
ValueType
10.1007/BF02492235DOI
Keywords:Programmed-temperature vaporization; Large-volume injection; Trichloroacetic acid; Thermal decarboxylation; Chloroform
Dewey Decimal Classification:500 Science > 540 Chemistry & allied sciences
600 Technology > 610 Medical sciences Medicine
Status:Published
Refereed:Yes, this version has been refereed
Created at the University of Regensburg:Yes
Item ID:34548