Multicenter Evaluation of Prosthesis Oversizing of the SAPIEN 3 Transcatheter Heart Valve. Impact on Device Failure and New Pacemaker Implantations

Pellegrini, Costanza ; Kim, Won-Keun ; Holzamer, Andreas ; Walther, Thomas ; Mayr, N. Patrick ; Michel, Jonathan ; Rheude, Tobias ; Nuñez, Julio ; Kasel, Albert M. ; Trenkwalder, Teresa ; Kaess, Bernhard M. ; Joner, Michael ; Kastrati, Adnan ; Schunkert, Heribert ; Hilker, Michael ; Möllmann, Helge ; Hengstenberg, Christian ; Husser, Oliver



Zusammenfassung

Introduction and objectives: A certain degree of prosthesis oversizing (OS) is recommended for the SAPIEN 3 to achieve device success. However, an increase in OS may increase permanent pacemaker implantation (PPI) rates. We therefore investigated the influence of OS on device failure and PPI. Methods: A total of 804 patients were treated with SAPIEN 3 at 3 centers. Multislice computed ...