Zusammenfassung
BACKGROUND: Ultrasound follow-up of transjugular intrahepatic portosystemic shunt (TIPS) is challenging due to the bent course of the stent-graft. OBJECTIVE: Aim of this retrospective study was to assess to which extent the combination of HR flow with Glazing Flow improves hemodynamic assessment in the ultrasound follow-up of TIPS. METHODS: Comparative studies with CCDS and High Resolution ...
Zusammenfassung
BACKGROUND: Ultrasound follow-up of transjugular intrahepatic portosystemic shunt (TIPS) is challenging due to the bent course of the stent-graft. OBJECTIVE: Aim of this retrospective study was to assess to which extent the combination of HR flow with Glazing Flow improves hemodynamic assessment in the ultrasound follow-up of TIPS. METHODS: Comparative studies with CCDS and High Resolution (HR)-Flow with Glazing Flow were evaluated regarding image quality and artifacts on a 5-point scale (0 = cannot be assessed up to 5 = maximum image quality without artifacts). In all cases, an experienced examiner performed the examinations with a 1-6MHz probe (Resona 7, Mindray). RESULTS: 61 ultrasound examinations in 48 patients were performed; the mean patient age was 54 +/- 14.2 years. The use of HR-Flow with Glazing Flow resulted in an improved flow display in 55/61 cases (90.2%). Both methods correlated well (r = 0.71), but HR flow with Glazing flow values were in general higher than CCDS values. The reading resulted in an average value of 2.52 +/- 0.54 for CCDS and 3.52 +/- 0.57 for HR flow with Glazing flow (p = 0.013). CONCLUSIONS: The combination of HR-Flow and Glazing Flow results in improved flow representation and reduction of artifacts in the ultrasound follow-up of TIPS.