Startseite UR

Preliminary study on the use of an inhomogeneous anthropomorphic Fricke gel phantom and 3D magnetic resonance dosimetry for verification of IMRT treatment plans

Gum, F ; Scherer, J ; Bogner, L ; Solleder, M ; Rhein, B ; Bock, M



Zusammenfassung

An inhomogeneous anthropomorphic phantom of the human thorax including lungs and spine was developed for verification of three-dimensional (3D) intensity-modulated radiotherapy (IMRT). The phantom and spinal cord were filled with undiluted Fricke gel, whereas the lungs were filled with a special low-density Fricke gel. Based on a computed tomography scan of the phantom. an intensity-modulated ...