Zusammenfassung
A comparative analysis of two different approaches developed to deal with molecular relaxation in photoacoustic spectroscopy is here reported. The first method employs a statistical analysis based on partial least squares regression, while the second method relies on the development of a digital twin of the photoacoustic sensor based on the theoretical modelling of the occurring relaxations. ...
Zusammenfassung
A comparative analysis of two different approaches developed to deal with molecular relaxation in photoacoustic spectroscopy is here reported. The first method employs a statistical analysis based on partial least squares regression, while the second method relies on the development of a digital twin of the photoacoustic sensor based on the theoretical modelling of the occurring relaxations. Methane detection within a gas matrix of synthetic air with variable humidity level is selected as case study. An interband cascade laser emitting at 3.345 mu m is used to target methane absorption features. Two methane concentration ranges are explored targeting different absorptions, one in the order of part-per-million and one in the order of percent, while water vapor absolute concentration was varied from 0.3 % up to 2 %. The results achieved employing the detection techniques demonstrated the possibility to efficiently retrieve the target gas concentrations with accuracy > 95 % even in the case of strong influence of relaxation effects.